Troubleshooting
Alarm
Alarm
priority
!!! 200
Rotary knob stuck
!
170
Sample line disconnected? Ventilation was started, but
!
170
Sample line occluded
!! 75
Second agent detected
!
100
Service date reached
!! 0
Speaker failure
!! 100
Synchronized ventilation
failure
180
Cause
Key is stuck or was pressed
for more than 10 seconds.
no exhaled CO
was
2
detected.
Sample line is occluded.
A second anesthetic agent
has been detected.
Maintenance is due.
The loudspeaker is faulty.
Inspiratory efforts of the
patient cannot be detected
due to faulty sensors.
Instructions for use Perseus A500 SW 1.1n
Remedy
Ventilation performance is not
affected.
Press and turn rotary knob
repeatedly. If alarm condition
persists, settings cannot be
adjusted anymore.
Disconnect the patient and
ventilate with a manual resus-
citator.
Switch off the device with the
device switch. Start manual
ventilation using "Add. O
Check sample line, water
trap, and patient-side filter.
Check sample line, water
trap, and patient-side filter.
Flush the system if necessary.
Check fresh-gas settings.
Wait for transition phase to
end.
Call DrägerService.
Use "ALARM RESET" to
acknowledge alarm.
Call DrägerService.
Use "ALARM RESET" to
downgrade alarm priority.
Check ventilation settings.
Change to a non-synchro-
nized ventilation mode.
Change to MAN/SPON mode
and ventilate manually.
".