Using The O 2 Flush; Using The Vaporizer - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Operation
Using the O
flush
2
The O
flush is used for flushing and quickly filling
2
the breathing system and breathing bag with O
while bypassing the vaporizer.
 Press the O
+ button. O
2
the button is held down.
Use of the O
flush can increase airway pressure
2
very quickly and abruptly change the gas
concentration.

Using the vaporizer

WARNING
Risk due to incorrect anesthetic agent
delivery
Filling the vaporizer with an incorrect
anesthetic agent results in an incorrect
anesthetic gas concentration.
Compare the color coding on the vaporizer
used to that of the anesthetic agent indicated
on the screen.
 Operate the vaporizer according to its
instructions for use.
78
2
flows for as long as
2
When the Vapor View option is installed, the
control dial and sight glass on the
Dräger-Vapor 3000 and D-Vapor 3000 are
illuminated:
Control dial position
T to <0
0
>0
WARNING
Risk due to faulty Vapor View option
If the Vapor View option is faulty, the
following information could be displayed
incorrectly on the screen: Anesthetic gas
type, filling level, control dial position,
prediction of anesthetic gas concentration
– Do not rely solely on the display on the
screen, but always pay attention to the
settings and values at the vaporizer as
well.
– Do not make therapy decisions based
solely on a displayed prediction.
WARNING
Risk of increased anesthetic agent delivery
Using the emergency O
may cause an increased anesthetic gas
concentration.
Carefully monitor the gas mixture.
Instructions for use Perseus A500 SW 1.1n
Illumination
Off
Medium
Bright
delivery (Add. O
)