Dräger Perseus A500 Instructions For Use Manual page 169

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Alarm
Alarm
priority
!
100
CO
sensor accuracy low
2
!! 135
Cooling fan failure
!! 140
CPAP changed to "Off"
!! 150
Emergency air inlet acti-
vated
!! 135
etCO
high
2
!! 135
etCO
low
2
!! 100
Exp. press. sensor failure
!
100
Expiratory flow sensor
expired
!
190
Expiratory flow sensor not
calibrated
Instructions for use Perseus A500 SW 1.1n
Cause
Accuracy of the CO
sensor
2
cannot currently be guaran-
teed.
An internal fan for evacuating
gases is faulty.
The set CPAP pressure could
not be achieved due to leak-
age. The system has
changed the CPAP setting to
"Off".
Negative pressure in the
breathing system due to
insufficient fresh gas in the
breathing bag. The system
utilizes ambient air to ensure
ventilation.
etCO
has exceeded the
2
upper alarm limit.
etCO
is below the lower
2
alarm limit.
Sensor calibration failed.
Accessory has been used too
long.
The sensor is not calibrated.
The breathing system has
been replaced or discon-
nected since last calibration.
Failure of the flow sensor.
Troubleshooting
Remedy
If the problem persists for
more than 10 minutes, call
DrägerService.
To prevent potential damage,
switch off system at your ear-
liest convenience. Increased
risk of fire.
Call DrägerService.
Check the breathing circuit for
tight connections and leak-
ages.
Increase fresh-gas flow.
Reapply the CPAP setting.
Use "ALARM RESET" to
acknowledge alarm.
Refill breathing bag (e.g., with
"O
+" or "Add. O
").
2
2
Increase fresh-gas flow.
Check the breathing circuit for
tight connections and leak-
ages.
Check ventilation.
Check ventilation.
Ensure that a suitable substi-
tute monitoring is available.
Perform the system test.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
Perform the breathing sys-
tem test.
Replace the flow sensor and
perform the breathing system
test.
169