Mounting The Vaporizers - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Handling O
gas cylinders
2
WARNING
Risk of explosion
When pressurized, O
combination with oil or grease.
Do not oil or grease the gas cylinder valve or
the pressure reducer of the O
not touch with oily or greasy fingers.
Gas cylinder valves must only be opened and
closed slowly. Do not use any tools with the screw
connection variants.
Have service personnel repair any leaky or stiff
gas cylinder valves.
WARNING
Risk due to failure of the O
Using a pressure reducer without the
required pressure sensor in place of a Dräger
pressure reducer or connecting a central
supply hose to the connection for the
external gas cylinders will prevent the
availability of the auxiliary supply from being
monitored during the system test and while
the system is in use. The back-up
functionality may be compromised.
If the O
gas cylinder monitoring function is
2
not available, the user must take measures to
ensure that the pressure indicator can be
read from the user's operating location. Do
not connect central supply hoses to the
external gas cylinder connection.
Instructions for use Perseus A500 SW 1.1n
is self-igniting in
2
cylinder. Do
2
gas supply
2

Mounting the vaporizers

Depending on the configuration, Perseus can be
operated with vaporizers which have a Dräger
Auto Exclusion plug-in adapter or a Selectatec
plug-in adapter. Dräger recommends using only
vaporizers that have been tested and are listed in
the list of accessories.
WARNING
Risk due to incorrect anesthetic agent
delivery
If the vaporizer is filled with the wrong
anesthetic agent or if it is not filled
sufficiently, incorrect anesthetic gas
concentrations or concentrations that are too
low can occur as a result.
Strictly observe the instructions for use of the
vaporizer. Compare the color coding on the
vaporizer with the anesthetic agent bottle.
The vaporizers used must conform to the
ISO 8835-4 or ISO 80601-2-13 standard. If the
internal patient gas measurement module fails, an
independent gas measurement system complying
with ISO/IEC 80601-2-55 must be used.
WARNING
Risk due to improperly mounted vaporizers
Incorrectly mounted vaporizers can cause
leakage. This can cause the fresh-gas flow to
be too low or contaminate the ambient air.
Patient and user can be endangered.
– Make sure that the connected vaporizers
are hanging vertically.
– When using D-Vapor vaporizers, make
sure that the power cable is not pinched.
– After mounting the vaporizers, perform a
leakage test.
Assembly and preparation
51