Fresh-Gas Delivery; Fresh-Gas Delivery (Electronically Controlled Gas Mixer) - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Technical data
Ambient conditions (continued)
Battery
For storage longer than 12 months
Maximum storage duration without
recharging
Ambient pressure
Relative humidity
CO
concentration
2
The permissible ambient conditions depend
on the accessories used. Observe related
instructions for use.

Fresh-gas delivery

All data are standardized to STPD conditions.
O
flush
2
O
flow for Aux. O
and Add. O
2
2
Range
Accuracy
Resolution of displayed value

Fresh-gas delivery (electronically controlled gas mixer)

All data are standardized to STAPD conditions.
O
concentration FG O
2
Fresh-gas flow FG flow
204
2
2
–15 °C to 40 °C
(5 °F to 104 °F)
–15 °C to 25 °C
(5 °F to 77 °F)
180 days
500 to 1060 hPa
(7.3 to 15.3 psi)
10 to 95 %, without condensation
Not relevant
25 to 75 L/min at 2.7 to 6.9 kPa x 100 (39 to
100 psi) supply pressure
Off; 2 to at least 10 L/min at 2.7 kPa x 100
(39 psi) supply pressure
±10 % of the set value for flows >2.0 L/min
1.0 L/min (up to 10 L/min)
5.0 L/min (above 10 L/min)
21 to 100 Vol% (carrier gas Air);
25 to 100 Vol% (carrier gas N
±5 % or ±2 Vol% (the larger value applies)
Off; 0.2 to 15 L/min;
±10 % or ±50 mL/min (the larger value applies)
Instructions for use Perseus A500 SW 1.1n