Patient-specific accessories and consumables
The patient-specific accessories and consumables must be removed from the device and, if necessary,
disassembled.
Reusable products:
If the reusable product has its own instructions for use, perform reprocessing in accordance with the
separate instructions for use. Further information can be found in the list of accessories.
If no separate instructions for use are available for the reusable product, perform reprocessing in ac-
cordance with the instructions in this IFU (see "Reprocessing of patient-specific accessories"
on page 257).
Disposable products:
Dispose of the disposable products.
WARNING
Disposable accessories (such as disposable electrodes) are for single-use only. Do not reuse
disposable accessories.
Removing the accessories
– ECG lead wire
– SpO2 intermediate cable
– Pouch M300 components
Instructions for use – Infinity
®
M300 and M300+ series – VG3.0
Reprocessing
251