Repairs - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Component
Flow sensors
AGS
AGS filter
Sample gas return filter
O-rings on the water trap
mount
Sealing, sieve insert,
and sintered filter on the
cylinder connection of
the external cylinder
pressure reducer
(optional)
Air filter mat
– Patient gas measure-
ment module
– Power supply unit
CLIC adapter (optional)
Diaphragm unit, valve
cone, discharge valve,
and safety valve spring
of the external cylinder
pressure reducer
(optional)
Lead-gel battery
(2 pieces)

Repairs

Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.
Instructions for use Perseus A500 SW 1.1n
Interval
If required, if calibration
is no longer possible, or
according to the config-
ured Infinity ID exchange
interval
If required, if filter is
soiled, or flow is no lon-
ger achieved
If required, if filter is
soiled, or flow is no lon-
ger achieved
Every 2 years
Every 2 years
Every 3 years
Every 3 years
Every 4 years
Every 6 years
Every 2 years
Or after determination of
remaining capacity
Task
Replace, see page 159
Replace, see page 161
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Maintenance
Responsible personnel
Users
Users
Service personnel
Service personnel
Service personnel
Experts
Service personnel
Users
Experts
Service personnel
Experts
199