Dräger Perseus A500 Instructions For Use Manual page 132

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Configuration
Vertical tab "Alarm config. 1"
Headline/
Parameter
General alarm behavior
Open "Quick setup"
if alarm occurs
"Second agent
detected" alarm
"xMAC low" alarm
"FiO
too high for
2
neonates" alarm
after [h:mm]
Priority of "Apnea
ventilation" alarm
Alarm behavior in "Pause" mode
Priority of "Pause
time expired" alarm
Default value for
"Timer" [mm:ss]
132
Setting range
On; Off
On; Off
On; Off
Off; 0:10 to 9:50
0:15
Medium; Low
Medium; Low
0:00 to 60:00
0:00 to 60:00
2:00
1:00
Description
Automatically open Quick setup
window in the event of an alarm.
Alarm when an anesthetic gas
mixture is detected.
Activate the xMAC low alarm.
Set the time until FiO
neonates alarm is sounded.
Specify the alarm priority when the
set minimum respiratory rate is not
reached in Pressure Support ven-
tilation mode.
Only available for electronically
controlled gas mixer.
Specify the alarm priority when the
set duration in the Pause mode
has expired.
0:00 to 60:00
Specify standard duration for
Pause.
0:30
Instructions for use Perseus A500 SW 1.1n