Operation
Mechanically controlled gas mixer:
WARNING
Risk of incorrect gas delivery
Use the flow values on the status display as
the primary source of information. Always
compare these values against the values on
the total flow tube.
Only use the total flow tube as the primary
source of information in the following
circumstances:
– The status display has failed.
– The values on the status display do not
match the values on the total flow tube.
 Use the flow control valves to adjust the
fresh-gas flows.
Use the total flow tube to check the total
flow set, see page 28.
Perseus is equipped with a minimum O
which ensures that a minimum quantity of oxygen
is delivered when N
O is used as the carrier gas,
2
see page 247. To prevent the soda lime from
drying out at an increased rate, the electronically
controlled gas mixer reduces the fresh-gas flow
FG flow automatically. This reduction takes place
depending on the set tidal volume VT and the set
respiratory rate RR.
CAUTION
Risk of patient injury
Unsuitable soda lime can result in disintegration
products from the anesthesia gases.
Use suitable soda lime such as Drägersorb Free.
CAUTION
Risk of patient injury
The use of minimum-flow or low-flow settings can
lead to accumulation of metabolic by-products in
the breathing system.
If minimum-flow or low-flow settings are used,
flush the breathing system regularly.
76
3 Set the ventilation settings (C).
4 To start therapy, press the rotary knob. A signal
tone is emitted when therapy starts.

Useful tips

Quick start with MAN / SPON mode
Electronically con-
trolled gas mixer
 In Standby mode,
touch Quick start
MAN/SPON.
Opening the start dialog in Standby mode
 Touch the screen in the monitoring area.
Or
delivery,
2
 Squeeze the breathing bag.
Starting when time is limited
When time is limited it is possible to bypass the
adjustment of the patient data and the ventilation
settings. Start the therapy as follows:
1 Touch the screen.
2 Check the displayed start values.
3 To start therapy, press the rotary knob.
4 Adjust the patient data and ventilation settings
as soon as possible.