After using the non-rebreathing system
1. Dismantle the non-rebreathing system and the sample line.
2.
Risk due to leakage from an open AGS inlet nozzle
► To prevent contamination of the ambient air with anesthetic gases, press the
Seal the inlet nozzle (4) on the AGS again with a sealing plug.
4.9

Connecting and replacing consumables

4.9.1
Disposable CO
► Connect or replace the CLIC absorber in accordance with its instructions for
use.
4.9.2
Reusable CO
As an alternative to disposable CO
be used.
4.9.2.1
Dismounting and emptying
1. Unscrew the CO
2. Remove and dispose of the disposable dust filter (2), if present.
3.
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
► Handle the soda lime carefully and do not spill it.
Empty out the used soda lime and dispose of it according to its instructions for
use.
|
Instructions for use
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
CAUTION
sealing plug back into the inlet nozzle after using a non-rebreathing system.
absorber
2
absorber
2
absorber (1) from the breathing system.
2
CAUTION
Assembly and preparation
absorbers, a reusable CO
2
absorber may also
2
75