Breathing System - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Ventilator (continued)
Pressure limitation Pmax
Lower pressure level Plow in APRV mode
Upper pressure level Phigh in APRV mode
Pressure amplitude above PEEP: ∆Psupp
In CPAP/Pressure Support and Pres-
sure Control - BIPAP / PS mode
In Volume Control - SIMV / AF / PS
mode
Positive end-expiratory pressure PEEP
Continuous positive pressure CPAP
Minimum pressure limit as per ISO 8835-5
and ISO 80601-2-13

Breathing system

Total volume including the disposable CO
absorber CLIC Absorber Free
In MAN / SPON mode
In automatic ventilation modes
Compliance (without breathing circuit)
In MAN / SPON mode
In automatic ventilation modes
Volume of absorber container
Reusable CO
absorber, filled
2
Instructions for use Perseus A500 SW 1.1n
PEEP +5 to 80 hPa (cmH
±10 % of the set value or ±3 hPa (cmH
larger value applies)
Off, 2 to 35 hPa (cmH
±10 % of the set value or ±2 hPa (cmH
larger value applies)
Plow +1 to 80 hPa (cmH
±10 % of the set value or ±2 hPa (cmH
larger value applies)
Off, 1 to (80 - PEEP) hPa (cmH
±10 % of the set value or ±2 hPa (cmH
larger value applies)
Off, 1 to (Pmax - PEEP) hPa (cmH
±10 % of the set value or ±2 hPa (cmH
larger value applies)
Off, 2 to 35 hPa (cmH
±10 % of the set value or ±2 hPa (cmH
larger value applies)
Off, 2 to 35 hPa (cmH
±10 % of the set value or ±2 hPa (cmH
larger value applies)
0 hPa (cmH
2
Typically 2.18 L
Typically 2.18 L
Typically 2.18 mL/hPa (mL/cmH
Typically 0.28 mL/hPa (mL/cmH
1.5 L
O);
2
O);
2
O);
2
O),
2
O);
2
O);
2
O);
2
O)
2
O)
2
Technical data
O) (the
2
O) (the
2
O) (the
2
O) (the
2
O);
2
O) (the
2
O) (the
2
O) (the
2
207