Functional Safety; Handling Infinity Id Components - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

WARNING
Risk due to electromagnetic fields
Although the medical device does not exceed
the applicable limiting values for
electromagnetic fields, such radiation can
interfere with the functioning of pacemakers.
All wearers of pacemakers should maintain a
distance of at least 25 cm (10 in) between
pacemaker and medical device.

Functional safety

The essential performance features comprise:
– Supplying the anesthesia workstation with O
If the O
supply (central gas supply or gas
2
cylinder) fails, an alarm is issued.
– Supply of the patient with adequately
oxygenated breathing gas
If the breathing gas contains insufficient levels
of O
, an alarm is issued.
2
– Patients are not supplied with excessively high
anesthetic gas concentrations
If excessively high anesthetic gas
concentrations are delivered, an alarm is
issued.
– Monitoring of the airway pressure and the
expiratory minute volume
Alarms are issued depending on the set alarm
limits.

Handling Infinity ID components

Ownership or purchase of this medical device with
RFID technology only includes the right to use the
medical device and RFID technology in
conjunction with products approved by Dräger and
in strict compliance with these instructions for use.
No intellectual property rights or any rights to the
use of the medical device or RFID technology are
hereby granted, either explicitly or implicitly, which
are contrary to the above-mentioned conditions.
Instructions for use Perseus A500 SW 1.1n
For your safety and that of your patients
2
11