Fresh-Gas Delivery (Mechanically Controlled Gas Mixer); Ventilator - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Fresh-gas delivery (mechanically controlled gas mixer)

All data are standardized to STAPD conditions.
O
concentration FG O
2
Setting range for fresh-gas flow (FG flow)
Electronic measurement of fresh-gas flow (FG
flow)
Range
Accuracy
Resolution of the value displayed on the
screen
Resolution of the value displayed in the sta-
tus display
Total flow tube
Range
Accuracy
Resolution of displayed value

Ventilator

Time-based settings
Respiratory rate RR
Respiratory rate during apnea ventilation
RRapn
Inspiratory time Ti
Maximum inspiratory time for supported
breaths (fixed setting)
Patient category "adult"
Instructions for use Perseus A500 SW 1.1n
2
21 to 100 Vol%
0 to at least 12 L/min volumetric per gas (O
and N
O)
2
0 to 15 L/min (O
, Air and N
2
±10 % or ±0.12 L/min (the larger value applies)
0.01 L/min (from 0 to 0.2 L/min)
0.02 L/min (from 0.2 to 0.5 L/min)
0.05 L/min (from 0.5 to 1 L/min)
0.1 L/min (from 1 to 15 L/min)
0.1 L/min (from 0 to 15 L/min)
All data are standardized to STPD conditions.
0 to 10 L/min
±10 % of the set value at 100 % O
flows >1 L/min
0.5 L/min
3 to 100 /min
±10 % of the set value or ±1 /min (the larger
value applies)
Off, 3 to 25 /min
±10 % of the set value or ±1 /min (the larger
value applies)
0.2 to 10 s
±10 % of the set value or ±100 ms (the larger
value applies)
4 s or 1 / (2 x RRapn) (the smaller value
applies)
Technical data
, Air
2
O)
2
and for
2
205