Patient-specific accessories and consumables
The patient-specific accessories and consumables
must be removed from the device and, if necessary,
disassembled.
Reusable products:
If the reusable product has its own instructions
for use, perform reprocessing in accordance
with the separate instructions for use.
If no separate instructions for use are available
for the reusable product, perform reprocessing
in accordance with these instructions for use.
Disposable products:
Dispose of the disposable products.
Removing the pneumatic medication nebulizer
When using a breathing circuit for adults:
B
1 Remove the nebulizer hose (A) from the
medication nebulizer (B) and from the nebulizer
port on the device.
2 Remove the medication nebulizer (B) from the
breathing circuit.
3 Disassemble the medication nebulizer in
accordance with the corresponding instructions
for use.
Instructions for use Savina 300 SW 5.n
A
When using a breathing circuit for pediatric
patients:
G
F
1 Remove the nebulizer hose (C) from the
medication nebulizer (D) and from the nebulizer
port on the device.
2 Remove the medication nebulizer (D) from the
breathing circuit.
3 Pull the catheter connector (E) out of the inlet
port.
4 Pull the adapter (F) out of the outlet port.
5 Remove the corrugated hose (G) from the
adapter (F).
6 Disassemble the medication nebulizer in
accordance with the corresponding instructions
for use.
Removing the CO
sensor
2
Prerequisites:
– The sensor plug at the rear of the device has
been removed.
A
B
1 Remove the CO
sensor (A) from the cuvette.
2
2 Remove the cuvette (B) from the patient port of
the Y-piece.
Reprocessing
E
C
D
175