Dräger Perseus A500 Instructions For Use Manual page 218

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Technical data
Operation characteristic values (continued)
Internal battery
Type
Capacity
Voltage
Fuse
Current
Backup time with new and fully charged bat-
tery
Minimum
Typical
Charging time (to reach full power)
Charging power
Gas supply
Gas quality
Oil content
Dew point
Particle size
Supply pressure for O
Maximum short term peak inlet flows at
6.9 kPa x 100 (or 100 psi) supply pressure
O
2
Air
Without suction unit
Including a directly connected Dräger
ejector suction unit
N
O
2
Drive gas
Gas supply connection
Gas cylinders (dimensions)
Diameter
218
, Air, N
O
2
2
Lead-gel battery
Sealed, maintenance-free
7.2 Ah
24 V
F15A 80V UL248
Maximum 15 A
30 minutes
150 minutes
At least 8 hours
Maximum 50 W
<0.1 mg/m3
5 °C (41 °F) at ambient temperature
Dust-free air (filtered with pore size <1 μm)
2.7 to 6.9 kPa x 100 (or 39 to 100 psi)
135 L/min
50 L/min
130 L/min
40 L/min
Not needed
Depending on configuration: DIN, NIST, DISS,
Air Liquide, SIS
100 to 140 mm (3.94 to 5.51 in)
Instructions for use Perseus A500 SW 1.1n