Leakages; Possible Leakage Causes - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Troubleshooting

Leakages

Leakages may result in the system not being
operational or being operational with limitations
only.
CAUTION
Risk due to contamination
Anesthetic gas can get into the ambient air as a
result of leakages.
Perform the leakage test before using the device.
Remedy all leakages.
CAUTION
Risk due to leakage at the valves
Leakages at valves can allow ambient air to enter
the breathing system and alter the composition of
the breathing gas.
Before using the device, perform the leakage test
to check the leak-tightness of all the valves.
WARNING
Risk of insufficient ventilation
Breathing gas may escape because of
leakages, with the result that the applied
volume is less than the set volume.
Perform the leakage test before using the
device. Remedy all leakages.
154

Possible leakage causes

– The CO
absorber or the CLIC adapter is not
2
securely screwed to the breathing system.
– The APL valve is not correctly fitted to the
breathing system or is not set to 30 hPa
(cmH
O).
2
– The breathing bag, the breathing hoses, the Y-
piece, or the microbial filter is incorrectly fitted
or damaged.
– The flexible arm for the breathing bag
(optional) is incorrectly fitted to the breathing
system. The sealing ring is soiled or damaged.
– The water trap is not connected.
– The sample line is not connected, is kinked, or
is leaking.
– The connections for the sample line are
damaged.
– The O-rings on the inspiratory port or
expiratory port are damaged, soiled, or
missing.
– The flow sensors are incorrectly installed or
damaged. The rear O-ring is missing.
– The upper part of the breathing system housing
is incorrectly fitted or damaged.
– The valves or seals of the breathing system are
damaged.
– The circuit plug is scratched or damaged.
– The filling or emptying connections on the
vaporizer are leaking or are open. The
vaporizer is incorrectly fitted. The O-ring is
missing or damaged. The control dial is not at
the 0 position.