Connecting The Gas Supply - Dräger Primus Infinity Empowered Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Connecting the gas supply

N
O AIR
2
1 Screw on the compressed gas hoses of the
central gas supply (CS) for O
front connections of the gas inlet block (A). The
two outer ports at the back are reserved for the
reserve gas cylinders.
A compressed air outlet for the optional en-
dotracheal suction system and an O
an external O
flow tube are optionally availa-
2
ble.
2 Connect the compressed gas hoses to the ter-
minal units.
3 Make sure all supplies are connected correctly
and functioning properly.
*
With the "Operation without nitrous oxide" option, con-
nection of an N
O gas supply is not possible.
2
Instructions for use Primus Infinity Empowered SW 4.5n
A
O
2
, AIR, N
O* to the
2
2
outlet for
2
CAUTION
Risk of supply failure
If all gas supplies (central and cylinder) are not
connected correctly, the reserve system will not
be available in the event of a gas supply failure.
Make sure all supplies are connected according
to the engraving on the gas inlet block and the il-
lustrations at the back of the machine. After con-
necting the supplies, ensure proper functionality.
CAUTION
Risk of device failure
In order for the anesthesia machine to operate as
specified, the supply pressures at the gas inlet
must be within a range of 2.7 to 6.9 kPa x 100.
Make sure this is the case before initiating opera-
tion.
4 All three LEDs (B) illuminate green.
The LEDs do not illuminate if the gas pressure
is <2.7 kPa x 100 or if the compressed gas
hose is not connected.
B
Assembly and preparation
59