Ensuring the gas supply

Perseus is equipped with an integrated anesthetic
gas receiving system (AGS).
1 Connect the suction hose to the nozzle (A) of
the receiving system.
2 Secure the suction hose (B) with strain
relievers.
3 Connect the probe of the suction hose to the
terminal unit of the scavenging system.
Instructions for use Perseus A500 SW 1.1n
A
B
As an option, the integrated anesthetic gas
receiving system can be operated in combination
with a control valve. Observe the assembly
instructions for the control valve.
CAUTION
Danger of ambient air contamination
If the anesthetic gas receiving system (AGS) is
not connected to the disposal system,
contamination of the ambient air with anesthetic
gas may result.
Make sure that the anesthetic gas receiving
system (AGS) and the disposal system are
correctly connected. The anesthetic gas
scavenging system must be functioning properly.
Assembly and preparation