Performing The Tests; Fully Automatic Tests - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Performing the tests

WARNING
Risk of patient injury
During the system test, the system is
pressurized.
To prevent patient injury, do not perform the
system test on the medical device if a patient
is connected.

Fully automatic tests

Prerequisites:
– Flex hoses are extended to their maximum
intended use length.
– O
supply is available.
2
1 Touch the button for the test.
Checklist
B
System test - do not connect patient.
2 Prepare the device according to the screen
display (A) and checklist (B), see page 106.
3 Start the test with the Start button (C).
Test progress and test results are indicated by the
progress display.
4 End the test with the OK button.
Instructions for use Perseus A500 SW 1.1n
5 After the system test, wait for 2 acoustic
NOTE
If the acoustic signals are not emitted, contact
service personnel.
Test interruption due to irregularities
If an irregularity is determined during a fully
automatic test, the test is stopped and information
regarding cause and remedy is displayed.
Test details
System test
A
Remedying the cause:
1 Remedy the cause of the interruption.
C
2 Touch the Repeat button (A) and repeat the
Accepting the irregularity:
 Touch the Accept button (B) and continue the
Accepted irregularities prevent the total result of
the system test from indicating "fully operational"
and are protocoled in the logbook.
signals.
If there is a leak, the leakage assistant (C) can
be used to support troubleshooting.
test of the component.
test.
System test
C
A
B
105