Dräger Perseus A500 Instructions For Use Manual page 165

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Alarm
Alarm
priority
!!! 220
Apnea (no CO
!! 255
Apnea (no CO
!!! 220
Apnea (no flow)
!! 255
Apnea (no flow)
!!! 220
Apnea (no pressure)
Instructions for use Perseus A500 SW 1.1n
Cause
)
No breathing or ventilation.
2
Sample line is not connected. Connect sample line to
)
No breathing or ventilation.
2
Sample line is not connected. Connect sample line to
No breathing or ventilation.
Fresh-gas flow is insufficient,
breathing bag is blocked or
positioned incorrectly.
The breathing hoses are
blocked or leaking.
No breathing or ventilation.
Fresh-gas flow is insufficient,
breathing bag is blocked or
positioned incorrectly.
The breathing hoses are
blocked or leaking.
No breathing or ventilation.
Fresh-gas flow is insufficient,
breathing bag is blocked or
positioned incorrectly.
The breathing hoses are
blocked or leaking.
Troubleshooting
Remedy
Start manual ventilation!
Check ventilation settings.
Check spontaneous breathing
ability of the patient.
breathing circuit.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breathing
ability of the patient.
breathing circuit.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breathing
ability of the patient.
Check fresh-gas settings and
position of breathing bag.
Check breathing circuit and
tube.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breathing
ability of the patient.
Check fresh-gas settings and
position of breathing bag.
Check breathing circuit and
tube.
Start manual ventilation!
Check ventilation settings.
Check alarm limit.
Check fresh-gas settings and
position of breathing bag.
Check breathing circuit and
tube.
165