Assembly and preparation

Selecting and connecting patient-specific accessories

Safety information

WARNING
Risk of increased rebreathing
Leakage between the inner and outer hose
cannot be detected when coaxial breathing
hoses are used.
To prevent insufficient gas exchange or CO
rebreathing, pay strict attention to the
measured gas concentration.

Connecting the breathing circuit and the filters

WARNING
Risk due to particles and dust
In order to protect the patient from particles
and dust, a filter must be used between the
inspiratory limb of the breathing system and
the patient.
Use a patient-side filter or a filter at the
inspiratory port.
WARNING
Risk of infection
If no microbial filter is used, the breathing
system may become contaminated with
disease-causing germs.
In this case, hygienically reprocess the
breathing system after each patient.
58
NOTE
Perseus is not made with natural rubber latex.
To minimize the risk of exposure to latex, use
latex-free breathing bags and breathing hoses.