Display of alarms

Alarms are signaled optically and acoustically.

Optical alarm signals

On the screen
In the event of an alarm, the system displays the
relevant alarm message in the alarm message
field (A). The parameter field (B) for the alarm-
generating parameter flashes.
C
B
B
B
Up to 8 alarms can be displayed simultaneously in
the alarm message field in the header bar (A). If
more alarms occur, the All alarms button (C) is
displayed in the header bar. Touching this button
opens the dialog window Alarms > Current
alarms with information about all active alarms,
see page 116.
On the Dräger Vapor 3000
When the Vapor View option is installed, alarms
concerning the vaporizer are signaled by blinking
illumination on the Dräger Vapor 3000 if their
cause can be remedied by adjusting or refilling the
vaporizer, e.g.:
– Inspiratory xMAC high
– Filling level of vaporizer low
– Vaporizer open
Instructions for use Perseus A500 SW 1.1n
A
B
B
B

Acoustic alarm signals

It always is the alarm with the highest priority that
is acoustically signaled. The signal is emitted until
either the cause of the alarm is remedied or the
Audio paused key is pressed.
Depending on the overall alarm situation, it is
possible that the 10-tone sequence for the alarm
priority "Warning" is only emitted as a 5-tone
sequence due to the coinciding occurrence of
alarms.
Regardless of the set alarm volume, the No O
delivery alarm is issued at maximum volume.
Alarms
2
115