Operation characteristic values (continued)
Drawer module (standard)
Drawer module (large)
Endotracheal suction with swivel arm and
accessories
Permissible total weight
Electromagnetic compatibility in compliance
with Directive 89/336/EEC
Protection classes
Device
Applied parts (connections for breathing
hoses)
Penetration of liquids
Classification in compliance with Directive
93/42/EEC, Annex IX
UMDNS Code Universal Medical Device
Nomenclature System - nomenclature system
for medical devices
GMDN Code Global Medical Device Nomencla-
ture - worldwide nomenclature for medical
devices
Use of latex

Device outlets

Serial ports
Protocol
Alarm delay time
Connector
Baud rate
Instructions for use Perseus A500 SW 1.1n
+ approximately 12 kg (26 lbs)
+ approximately 16 kg (35 lbs)
+ approximately 6 kg (13 lbs)
approx. 330 kg (728 lbs)
Tested in compliance with IEC 60601-1-2
I, in compliance with IEC 60601-1
TYPE BF
IP20 in compliance with IEC 60529,
fulfills IEC 60601-2-13
II b
10-134
37710
Perseus is not made with natural rubber latex.
COM 1 and COM 2;
Only connect devices that meet the require-
ments of IEC 60950-1 for ungrounded SELV cir-
cuits and the requirements of IEC 60601-1 (as
of the 2nd edition) for touchable secondary cir-
cuits with a maximum nominal voltage of 24
VDC.
MEDIBUS.X
Typically <2 s
9-way Sub-D
1200, 2400, 4800, 9600, 19200, 38400 baud
Technical data
221