Admitting a patient

You can admit a patient at the M540 manually by
entering the demographic data in the Patient setup
tab.
You can also admit a patient over the network by
pulling the data from an HL7/ADT interface. This is
either possible if the M540 is part of an IACS con-
figuration that is connected to the Infinity network,
the patient data can be retrieved over the network
and transferred to the M540.
WARNING
Monitors in a care area may seem identical but
may use different default alarm settings be-
cause of different profile assignments. After
admitting a patient, always verify that the set
alarm limits are appropriate for the patient.
To admit the patient manually
1 Press the Menu function key.
2 Touch the Patient setup tab.
3 Touch Patient category and then touch the ap-
propriate category (Adult, Pediatric,
Neonate). The message Changing category
will change alarm settings and algorithmic
processing appears. For more details on the
categories, see page 81.
4 Touch OK. The Patient setup dialog window
closes.
5 Press the Menu function key.
6 Touch the Patient setup tab.
7 Touch Name and use the onscreen keyboard to
enter the name (up to 25 characters)
8 Touch Confirm.
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
9 Touch ID and use the onscreen keyboard to en-
ter the ID number (up to 12 characters) and
then touch Confirm.
10 Touch Admit date, then touch each of the fol-
lowing: Day, Month, and Year to enter the ap-
propriate date.
11 Touch OK to confirm the data entry.
Use the arrows to scroll up or down to change
the data. To scroll faster through the data (for
example, the year), touch and hold the arrow.
12 Touch Birth date, then touch each of the follow-
ing: Day, Month, and Year to enter the appro-
priate date.
Use the arrows to scroll up or down to change
the data. To scroll faster through the data (for
example, the year), touch and hold the arrow.
13 Touch OK to confirm the data entry.
14 Touch Physician and use the onscreen key-
board for entering the name of the physician (up
to 12 characters).
15 Touch Confirm to confirm the data entry.
NOTE
The height and weight of the patient must be en-
tered at the Cockpit when the M540 is docked in
an IACS configuration. When the M540 is on wire-
less transport or in standalone mode, the height
and weight can be entered at the ICS.