Dräger Infinity Acute Care System M540 Instructions For Use Manual page 174

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

SpO
and Pulse CO-Ox monitoring with Masimo SET MCable
2
Selection
Available settings
Sensitivity
– Normal (default) – standard mode
1)
mode
– APOD (adaptive probe off detection) –
– Max. – provides maximum sensitivity for
1)
Fast SAT mode
On, Off (default)
When the Averaging time setting is set to 2
to 4 s or 4 to 6 s, the Fast SAT mode selec-
tion is ghosted.
1)
Color
Red, White (default), Yellow, Green, Light
blue, Blue, Purple, Orange
Change parame-
A list of currently available parameters.
ter
1)
This setting is a patient default which may be unique for each patient category; it is part of the profile.
NOTE
The password-protected alarm setting SpO
For more detailed information see page 105.
174
the least sensitive mode for detecting a
reading on patients with low perfusion.
Provides the best detection for detached
sensors. This mode is useful for patients
at particular risk for sensors becoming
detached such as children or patients
who are restless.
poor signals (this mode is recommended
for patients with low perfusion or when
the low perfusion or low signal quality
message is displayed in APOD or
Normal sensitivity mode. Max. mode is
not recommended for care areas where
patients are not monitored visually, such
as general wards. It is designed to inter-
pret and display data at the measuring
site when the signal may be weak due to
decreased perfusion.)
sensor off provides additional SpO
2
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Description
Determines the level of detection sen-
sitivity.
Activates rapid tracking of arterial ox-
ygen saturation changes.
Determines the color of the wave-
forms, and the parameter labels and
values.
Changes the parameter box to a dif-
ferent parameter.
alarm configuration.
2