Troubleshooting
CO
2
Priority
Message
!!
etCO
> (alarm limit)
2
etCO
< (alarm limit)
2
(except inCO
!
CO
check airway
2
1)
adapter
!
CO
H/W failure
2
!
CO
incompatible
2
1)
sensor
1)
In the parameter box the parameter value is replaced by ***
282
Cause
The parameter value is
above/below the set up-
per/lower alarm limits.
)
2
– The mainstream sensor
is not properly seated on
the adapter
– There are secretions in
the adapter.
– There is sensor zero drift
1)
CO
sensor hardware fail-
2
ure.
– The M540 has detected
that the used main-
stream sensor is not
compatible with the se-
lected sensor type set-
ting (reusable/dispos-
able)
– Secretions in the adapt-
er
– Sensor zero drift
– High inspiratory CO
concentration
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Remedy
– Check the patient and treat
if necessary.
– Change the alarm limits.
– Make sure the mainstream
sensor is attached properly
to the adapter.
– If message persists, clean
or replace the airway adapt-
er.
– If message persists though
the airway adapter is clean,
zero the sensor.
Contact your technical per-
sonnel.
– Use the airway adapter
type the system is config-
ured for or adjust the airway
adapter setting (see
page 249).
– If the message persists,
clean or replace the airway
adapter.
– If the message persists
even though the correct air-
way adapter type is select-
2
ed and the airway adapter
is clean, zero the sensor.
– If the message still persists,
the inspiratory CO
might not be accurate.
Check the patient and ven-
tilation.
value
2