Dräger Infinity Acute Care System M540 Instructions For Use Manual page 275

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Priority
Message
SpO2 only mode
Masimo
rainbow
SET only
Parameter values are displayed
for SpO
, PLS, and PI; the Masi-
2
mo rainbow SET parameter val-
ues are replaced by ***
None
SpO
searching
2
Any SpO
2
MCable
!
SpO
sensor failure
2
Any
SpO
2
MCable
!
SpO
sensor off
2
Any Masi-
mo MCa-
ble
!
SpO
sensor unplugged
2
In the parameter box the param-
eter value is replaced by the fol-
lowing depending on which MCa-
ble is used: *** for the parameters
SpO2, PLS,
SpHb/SpHbv, PVI, SpCO, SpOC,
SpMet; blanks for the parameter
PI.
3)
In the parameter box the parameter value is replaced by ***
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Cause
The device cannot calibrate
the Masimo rainbow SET pa-
rameters and is attempting
to display the standard Masi-
mo parameters.
3)
The sensor is searching for
valid pulses to compute a
measurement value.
3)
The MCable has detected a
hardware failure with the
SpO
3)
The Masimo MCable has de-
tected that the SpO
is no longer attached to the
patient.
The SpO
ble or sensor is unplugged.
sensor.
2
sensor
2
intermediate ca-
2
Troubleshooting
Remedy
Remove and reapply
the sensor. If the prob-
lem persists, contact
your technical person-
nel.
Verify proper sensor
application.
– Make sure the
SpO
sensor is
2
properly attached to
the patient and all
cables are properly
connected.
– Replace sensor.
– Contact Dräger
Technical support.
Reattach the SpO
2
sensor.
– Verify that the cable
and the sensor are
properly connected.
– Check for defective
sensor.
275