Cleaning and disinfection

Overview of cleaning and disinfecting the M540 and its accessories

Clean and disinfect the device or device parts be-
fore each maintenance step – and also when re-
turning for repair.
Continuous exposure to moisture can damage the
M540 and its peripheral devices. Please read the
following instructions carefully before cleaning any
device.
– Do not spray cleaning agents on the M540 or
peripheral devices. Wipe them with a cloth
moistened with soapy water.
– Disinfect the surfaces with a gauze moistened
with one of the approved agents (see
page 293).
– Dry thoroughly with a lint-free cloth.
Before cleaning any device, read the general safety
precautions under "General safety information" on
page 9.
292
Instructions for use Infinity Acute Care System - Infinity M540 SW VG4.n

Cleaning and disinfecting precautions

WARNING
Because of the danger of electric shock, never
remove the cover of any device while it is in
operation or connected to power.
WARNING
Do not immerse or rinse the device and its pe-
ripherals. If you spill liquid on the device (in-
cluding the battery or accessories), or acci-
dentally immerse it in liquid, disconnect the
device from the power source and allow it to
dry completely for at least 24 to 48 hours. Con-
tact your technical personnel regarding the
continued safety of the device and its periph-
erals before placing it back in operation.
CAUTION
To avoid damaging the device, do not use sharp
tools or abrasives. Never immerse electrical con-
nectors in water or other liquids.
CAUTION
Do not autoclave accessories.
CAUTION
Never immerse electrical connectors or the NIBP
connector.
Refer to the following instructions for use for the fol-
lowing devices:
For cleaning instructions regarding devices that
connect to the IACS, refer to the instructions for use
entitled Infinity Accute Care System -- Monitoring
Applications (Software VG4.n)
For specific cleaning instructions regarding the
Cockpit, refer to the instructions for use entitled In-
finity Acute Care System - Infinity Medical Cockpits.