Alarms
Zeroing an individual invasive blood pressure from
the M540 (see page 215) has the following effects:
– The invasive blood pressure limit alarm and
static alarm for that parameter are suppressed
from the time the button is pressed until
30 seconds after the zeroing procedure is com-
pleted.
– If the zeroed parameter is ICP or ART, the CPP
limit alarm is also suppressed from the time the
button is pressed until 30 seconds after the ze-
roing procedure is completed.
The following alarm conditions cancel the suppres-
sion of alarms caused by zeroing invasive blood
pressures:
– Invasive blood pressure parameter is outside of
the measuring range (too high/low)
– Invasive blood pressure transducer failure due
to a hardware failure, such as an open circuit or
short circuit within the cable.
– Unplugged transducers
– Disconnected hemodynamic pods
– A wedge pressure measurement that ends be-
fore the 30-second zeroing period ends, acti-
vates the alarm limit for the parameter PA M
only
92
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n

Privacy mode

Privacy mode is available when the M540 is docked
in an IACS configuration, on wireless transport, or
in standalone mode provided the patient is admitted
at the ICS. When privacy mode is activated, the fol-
lowing happens at the M540:
– All patient data are removed from the M540, but
continues to display at the ICS
– The alarm bar is deactivated
– Acoustic alarm signals are only provided at the
ICS
– The M540 displays Privacy, Touch screen to
resume monitoring
You can activate privacy mode only if the patient is
also admitted at the ICS. To activate or deactivate
this feature, see page 66.

Standby mode

When Standby mode is activated, the following
happens at the M540:
– All patient data are removed from the screen
– All monitoring (including acoustic and visual
alarm signals) is suppressed
– Active alarms are considered acknowledged by
the user
– All recordings are canceled at the Cockpit (pro-
vided the M540 is docked in an IACS configura-
tion)
– The M540 displays Standby, Touch screen to
resume monitoring