Alarms
Pre-silencing alarms
This function allows you to pre-silence (audio
pause in advance) potential alarm conditions be-
fore they occur. Pre-silencing allows you to concen-
trate on a procedure without being interrupted by
continuous acoustic alarm signals arising from po-
tential alarm conditions. A pre-silence period lasts
two minutes.
Pre-silencing alarms has the following effect:
– Any alarm conditions are reported visually by a
corresponding alarm message and a blinking
parameter box (see page 87).
– A single alarm tone sequence is generated for
the first occurrence of an alarm condition of low,
medium or high priority.
– The alarm message ? Audio paused appears
in the far right field of the header bar along with
a countdown timer and the following symbol:
– If multiple alarm conditions arise during an ac-
tive pre-silence period, the M540 triggers a sin-
gle alarm tone sequence for the highest-grade
alarm event.
NOTE
Pre-silencing alarms is not possible when the qui-
et mode is deactivated.
Initiating a pre-silence period
You can initiate a pre-silence in several ways:
– From an M540 in standalone mode or on wire-
less transport
– From an ICS
– From the remote view of another Infinity monitor
within the same monitoring unit
– From the Cockpit when the M540 is docked in
an IACS configuration
94
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
To initiate a pre-silence period remotely is only pos-
sible when the remote control setting of the remote
device is also activated. Refer to the corresponding
instructions for use for information on how to acti-
vate the remote control feature.
To pre-silence alarms at the M540
Press the yellow
The pre-silence state can be cancelled by pressing
the same key again that initiated the pre-silence pe-
riod. All alarm events are again reported as usual.
To pre-silence alarms remotely
Press the yellow
bar of the ICS to pre-silence alarms for all as-
signed patients. Press the same button in the
viewport area to pause alarm tones for an indi-
vidual patient. For more information, refer to the
ICS instructions for use.
Refer to the instructions for use of any remote
device within the same monitoring unit for in-
structions on how to initiate an audio pause.
The pre-silence state can be cancelled by pressing
the same key again that initiated the pre-silence pe-
riod. All alarm events are again reported as usual.
key on the M540.
key on the main menu