Impedance respiration (RRi)

Respiration parameter setup functions

All respiration parameter setup functions take place
in the respiration dialog window (see "Accessing
the respiration dialog window" on page 123).
Selection
1)
Resp. lead
Relearn
1)
Mode
1)
Marker
1)
Monitoring
1)
Apnea time
1)
Apnea archive
1)
This setting is a patient default which may be unique for each patient category; it is part of the profile.
1)
Coincidence
124
Available settings
I, II (default)
None
Auto (default), Manual
On, Off (default)
– On (default in neonatal mode)
– Off (default in adult/pediatric mode)
Off, 10, 15 (default), 20, 25, 30 s
– Off
– Str/Rec – a recording as well as an event
storage is triggered automatically in re-
sponse to an apnea.
– Store (default) – a waveform segment is
stored in response to an apnea.
– Record – a recording is triggered automat-
ically in response to an apnea.
On, Off (default)
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
The limits dialog window contains the Auto set and
Alarm buttons for configuring the alarm functions.
For detailed alarm setup information, see "Config-
uring a patient's alarm settings" on page 98.
Settings
Description
Selects the lead for respiration
monitoring.
Initiates a relearning of the respi-
ration signal.
Determines the processing
mode for the breath-related im-
pedance change.
Superimposes a vertical line on
the respiration waveform when a
breath is detected (see
page 122).
Turns respiration monitoring on
or off.
Determines how long an apnea
has to last before an alarm is trig-
gered.
Determines what happens in re-
sponse to an apnea.
Determines whether you are
alerted when the respiratory rate
is within 20 % of the heart rate,
which is an indication that the
M540 is counting heart beats as
respiration.