Rainbow Set - Dräger Infinity Acute Care System M540 Instructions For Use Manual

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Technical data
Pulse Oximetry (SpO

rainbow SET

Adult and pediatric sensors
Neonatal sensors
Parameter display
Measuring method
Measurement range
(Infinity MCable – Masimo SET)
Measurement range
(Infinity MCable – Masimo rainbow SET)
Resolution
(Infinity MCable – Masimo SET)
Resolution
(Infinity MCable – Masimo rainbow SET)
Maximum update interval
Accuracy
(Infinity MCable – Masimo rainbow SET)
PI accuracy
328
) Infinity MCable – Masimo SET and Infinity MCable – Masimo
2
LNCS DCI, LNCS DCIP, LNCS TC-I, LNCS TF-I, LNCS YI,
LNCS Adtx, LNCS Pdtx, LNCS Adtx-3, LNCS Pdtx-3
LNCS Inf, LNCS Inf-3, LNCS Neo, LNCS Neo-3, LNCS
NeoPt-3, LNCS NeoPt, LNCS YI, LNCS SofTouch, Neo-Pt-
500
Masimo SET MCable: Pulse oximetry (SpO
(PLS), perfusion index (PI)
Masimo rainbow SET MCable: Pulse oximetry (SpO
rate (PLS), perfusion index (PI), SpHb (total hemoglobin),
SpOC (total oxygen saturation), SpCO (carbon monoxide in
hemoglobin), SpMet (methemoglobin saturation), PVI (pleth
variability index)
Absorption-spectrophotometry
SpO
2
PLS: 26 to 239 bpm
PI: 0.00 to 20 %
SpHb/SpHbv: 0.0 to 25.0 g/dL (0.0 to 15.5 mmol/L)
SpOC: 0 to 35 mL/dL
PVI: 0 to 100 %
SpCO: 0 to 99 %
SpMet: 0 to 99.9 %
SpO
2
PLS: 1 bpm
PI: 0.01 %
SpHb/SpHbv: 0.1 g/dL (0.1 mmol/L)
SpOC: 1 mL/dL
PVI: 1 %
SpCO: 1 %
SpMet: 0.1 %
30 s
SpHb / SpHbv for 8 to 17 g/dL: ±1 g/dL
SpCO accuracy for 1 to 40 %: ±3 %
SpMet accuracy for 1 to 15 %: ±1 %
±10 %
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
: 1 to 100 %
: 1 %
), pulse rate