ECG, arrhythmia, and ST segment
ST complex dialog windows
The number of displayed ST complexes depends
on the connected lead wire set. You can view all ST
complexes or zoom in on a single ST complex to
view it in greater detail.
The following functions are available in the single
ST complex dialog window:
– Changing the ISO point
– Changing the ST measuring point
To access the general ST complex dialog
window
1 Touch the ST parameter box.
2 Touch the Settings tab.
3 Touch ST complex.
156
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Zooming in on a single ST complex
The following diagram shows a single ST complex
when you zoom in on one ST complex.
F
A
E
C
A ST label (unique for each ST lead)
B Single complex window
C When ISO: or ST: button is selected, left and
right arrows allow you to move the ISO/ST mea-
suring point
D Confirm button
E ST: button
F ISO: button
To zoom in on a single ST complex
1 Touch the ST parameter box.
2 Touch the Settings tab.
3 Touch ST complex. The general ST complexes
dialog window displays.
4 Touch a single ST complex window.
C
B
D