Cleaning and disinfection

Cleaning and disinfecting the M540 and M500

Use only the approved cleaning agents listed on
page 293, unless otherwise specified.
M540, M500, and power supply
precaution
WARNING
Do not immerse or rinse the device and its pe-
ripherals. If you spill liquid on the device (in-
cluding the battery or accessories), or acci-
dentally immerse it in liquid, disconnect the
device from the power source and allow it to
dry completely for at least 24 to 48 hours. Con-
tact your technical personnel regarding the
continued safety of the device and its periph-
erals before placing it back in operation.
To clean the M540 and M500
1 Clean the M540/M500 with a gauze pad moist-
ened with soapy water or with an approved
cleaning agent.
2 Dry thoroughly with a lint-free cloth.
294
Instructions for use Infinity Acute Care System - Infinity M540 SW VG4.n
To disinfect the M540 and M500
1 Disinfect the M540/M500 with a gauze pad us-
ing an approved cleaning agent.
2 Dry thoroughly with a lint-free cloth.
CAUTION
Do not steam autoclave, gas sterilize, or immerse
the M540 in liquid or cleaning solutions. Do not
subject the M540 to intense vacuum.
To clean the power supply
Clean the outside enclosure of the power supply
with a cloth using an approved cleaning agent.