Configuring general settings
This section describes the setup functions of the
Main dialog window.
To access the Main dialog window
1 Press the Menu function key.
2 Touch the Main tab.
Selection
All alarms pause
or
All alarms off
Show all leads
Label IBP
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Main dialog window
Available settings
None
None
Unlabeled pressure labels:
Pod 1A label, Pod 1B label,
Pod 1C label, Pod 1D label
Pressure labels:
GP2, ART, LV, LA, PA, CVP,
ICP, RA, RV
Generic labels:
GP1 through GP4
Description
When the All alarms pause button is select-
ed, all alarm functions are temporarily sup-
pressed for a selected time. The alarm func-
tion is automatically activated when the alarm
pause timer times out.
When the All alarms off button is selected, all
alarm functions are suppressed until you se-
lect the button again to activate the alarm func-
tion.
To configure this button, see page 237.
Displays waveforms for all ECG leads.
Assigns a label name to the IBP labels.
System configuration
237