Battery recharging time
Mode of operation
Environment
Humidity (non condensing)
Temperature
Atmospheric pressure
Drop
IEC 60068-2-32: 1975 +A1: 1982, +A2:
1990, Procedure 1
Transportation
Risk management
Protection against liquid ingress
Communications
Wired network
Wireless network
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
100 % capacity: approximately 6.5 hours for a completely dis-
charged battery
90% capacity: approximately 5.5 hours for a completely dis-
charged battery
70 % capacity: approximately 4 hours for completely dis-
charged battery
Continuous (with power coupling through the M500)
Operating: 15 to 85%
Storage: 10 to 95 %
Operating: 10 to 40 °C (50 to 104 °F)
NOTE: At ambient temperatures above 35 °C (95 °F) the bat-
tery may not be charging even while it is docked in the M500;
however, the battery icon still indicates the actual battery
charge.
Storage: –20 to +60 °C (–4 to +140 °F)
Operating: 485 to 795 mmHg (64.7 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
Drop once on each of six surfaces from a height of 1 m (3.2 ft)
Per International Safe Transit Association (ISTA)
IPX4 (protected against splashing water) per IEC 60529
802.3 100BaseT Ethernet when connected to the M500.
Optically isolated connection between the M540 and M500.
Complies with IEEE 802.11b/g WLAN standards. Supports
WPA2 security.
Technical data