Invasive Blood Pressure (Ibp); Non-Invasive Blood Pressure (Nibp) - Dräger Infinity Acute Care System M540 Instructions For Use Manual

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Technical data
Detection threshold
Measurement range
Resolution
Measuring accuracy
Apnea detection intervals

Invasive blood pressure (IBP)

Measuring method
Resolution
Measurement range
Dynamic range
Zero balance range
Filter settings
Accuracy
IBP update interval
Response time
(at 90 % of pressure change)
Transducer specifications

Non-invasive blood pressure (NIBP)

324
Manual mode: 0.15 to 2.0 Ω
Auto mode: 0.2 to 1.5 Ω
0 to 155 bpm
1 bpm
±1 or 2 % of rate (whichever is greater)
Off, 10, 15, 20, 25, and 30 s
Resistive strain gauge transducer
1 mmHg (0.1 kPa)
–50 to 400 mmHg (–6.6 to +53.3 kPa)
(GP1 to 4, ART, PA, wedge pressure, CVP, LA, LV, RV, RA,
ICP)
Before zeroing: –250 to +600 mmHg
(–33.3 to +79.9 kPa)
After zeroing: –50 to +400 mmHg
(–6.6 to +53.3 kPa)
±200 mmHg (±26.6 kPa)
User selectable DC to 8 Hz, DC to 16 Hz
±1 mmHg or ±3 % (whichever is greater)
excluding the transducer
4 s
14 beats + 2 s (ART, LV, GP1, GP2, GP3, GP4)
8 beats + 2 s (PA, RV)
16 s (CVP, RA, LA, ICP)
Transducers with a resistance of 200 Ω to 3000 Ω and an
equivalent pressure sensitivity of 5 μV/V/mmHg ±10 %
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n