Connecting the lead wires for neonatal ECG monitoring

The ECG lead wire sets connect directly to the
M540:
D
A M540 ECG connector
B Neonatal ECG adapter cable
C Neonatal ECG electrodes
D Spacer
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
A
B
C
ECG, arrhythmia, and ST segment
To connect the ECG lead wire set
1 Insert the neonatal ECG adapter cable (B) into
the recessed ECG connector (A) on the side of
the M540.
Orient the neonatal ECG adapter cable (B) so
the exposed pins face toward you as you push
it firmly into the channel.
NOTE
An ECG lead wire set can rest in the ECG connec-
tor of the M540 without actually being connected.
Make sure that all ECG lead wire sets are pushed
firmly into the ECG connector of the M540.
2 Insert the spacer (D) to protect the unused ECG
lead pins on the M540.
3 Connect the individual neonatal ECG elec-
trodes (C) to the neonatal ECG adapter cable
(B).
For information on applying the electrodes to the
patient, refer to the illustrations starting on
page 137.