Password-Protected Masimo Rainbow Set Setup Functions; Set Setup Functions - Dräger Infinity Acute Care System M540 Instructions For Use Manual

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Selection
Available settings
Change parame-
A list of currently available parameters.
ter
1)
This setting is a patient default which may be unique for each patient category; it is part of the profile.
2)
If the venous blood source was selected for SpHb Cal, the parameter label changes from SpHb (arterial
blood source) to SpHbv.

Password-protected Masimo rainbow SET setup functions

Additional Masimo rainbow SET setup functions
take place in the Settings 2 page which is protect-
ed by a clinical password.
Selection
1)
SpHb Cal
1)
PVI Averaging
1)
This setting is a patient default which may be unique for each patient category; it is part of the profile.
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
SpO
Available settings
– Arterial (default)
– Venous
– Short
– Long (default)
and Pulse CO-Ox monitoring with Masimo SET MCable
2
Description
Changes the parameter box to a dif-
ferent parameter.
Description
Selects the blood sampling
source which is used to calcu-
late the SpHb value.
The SpHb value changes to
SpHbv when the SpHb Cal set-
ting Venous is selected.
Determines how responsive the
monitor is to rapid physiological
changes while tracking pleth
variability index.
A longer averaging time pro-
vides a more accurate result.
However, in clinical situations
where rapid physiological
changes have to be monitored,
use a shorter averaging time.