Dräger Infinity Acute Care System M540 Instructions For Use Manual page 238

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

System configuration
Selection
Rest ECG report
Standby
Privacy mode
Discharge
M540 Audio
238
Main dialog window
Available settings
None
None
None
– Cancel
– Discharge
– On
– Off (default
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Description
Prints a 12-lead ECG report for an adult or pe-
diatric patient during 12-lead monitoring when
the M540 is in wireless mode.
When the M540 is docked, the feature is deac-
tivated. In this case a 12-lead ECG report can
only be printed from the Cockpit.
The report can be configured to include the fol-
lowing demographic data (refer to page 239):
– Gender
– Race
If the ECG filter is set to ESU, the report can-
not be generated.
Enters standby mode.
Enters privacy mode. This mode is only avail-
able when the M540 is docked in an IACS con-
figuration or when it is on wireless transport
and the patient is admitted at the ICS.
If the patient is not admitted at the ICS, the but-
ton is displayed but it is not activated. For more
information, see page 66.
Discharges a patient and deletes all the pa-
tient's data.
When the M540 is initially docked to a Cockpit,
the M540 Audio setting is Off and the speaker
off icon displays. This setting can be turned On
and if needed, the alarm volume and pulse
tone volume can also be adjusted.
When the M540 is on transport, in wireless
mode or in standalone configurations, this set-
ting is always activated.