Recommended Separation Distances - Dräger Infinity Acute Care System M540 Instructions For Use Manual

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Technical data
NOTE
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans-
mitters, an electromagnetic site survey should be considered. If the measured field strength in the lo-
cation in which the equipment is used exceeds the applicable RF compliance level above, the equip-
ment should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the equipment.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances

Maximum P
EIRP
(Watts)
0.001
0.003
0.010
0.040
0.100
0.200
0.250
1.000
2.000
3.000
10.00
100.00
1)
Information regarding separation distances (IEC 60601-1-2:2007, tables 4 and 6)
340
150 kHz to 800 MHz
1
Distance
(m)
0.04
0.06
0.12
0.21
0.38
0.54
0.6
1.2
1.7
2.08
3.8
12
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
800 MHz to 2.5 GHz
1
Distance
(m)
0.07
0.12
0.23
0.4
For example: WLAN 5250
0.73
For example: WLAN 2440
(Europe), Bluetooth
1.03
For example: WLAN 5250
(Europe)
1.03
For example: DECT-devices
2.3
For example: GSM 1800 /
GSM 1900 / UMTS mobile
phones, WLAN 5600 (not in
Europe)
3.25
For example: GSM 900 mo-
bile phones
3.98
7.27
23
Comments (if applicable)