M540 Battery Messages; Banners - Dräger Infinity Acute Care System M540 Instructions For Use Manual

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Troubleshooting

M540 battery messages

Priority
Message
!
Low battery
!!
Recharge battery

Banners

Priority
Message
!!!
All alarms off
!!!
All alarms paused with
countdown timer
!!
All alarms paused with
countdown timer
264
Cause
The M540 is undocked and
the battery charge has
reached 10% of the remain-
ing charge.
An advisory tone sounds
every 20 seconds.
The M540 is undocked and
the battery charge has
reached 5% of the remain-
ing charge.
Cause
The All alarms off func-
tion is set to No timeout
(see page 255) and you
select the All alarms off
button.
The All alarms paused
function is set to a time
(see page 255) and you
select the All alarms
paused button.
The yellow All alarms
paused key was pressed.
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Remedy
Return the M540 to the M500 to
recharge the battery.
For information on how to config-
ure the priority of the alarm tone
that accompanies this message,
see page 257.
Remedy
Select the All alarms off but-
ton again to remove the ban-
ner.
Select the All alarms
paused button again to re-
move the banner.
Press the key again to re-
move the banner.