Pausing Alarm Monitoring Temporarily - Dräger Infinity Acute Care System M540 Instructions For Use Manual

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Pausing alarm monitoring temporarily

If the password-protected alarm pause feature is
activated (see page 241), you can pause alarm
monitoring temporarily. The alarm pause duration is
adjustable from 1 minute to 5 minutes.
NOTE
If the French NFC mode is activated (see
page 249), you cannot pause alarm monitoring for
more than 3 minutes.
The following happens when you pause alarm
monitoring:
– Acoustic and visual alarm signals for new alarm
conditions are suppressed for all parameters
until alarm monitoring begins again
– Alarm signals for any active alarm condition
stop immediately
– The alarming parameter box and alarm bar re-
turn to the pre-alarm state
– Alarm messages are removed from the alarm
message field in the header bar
– The far right field of the header bar turns yellow
and displays the alarm message All alarms
pause, a countdown timer, and the following
symbol:
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
To pause alarm monitoring temporarily
1 Press the Alarms function key.
2 Touch All alarms pause.
As soon as the alarm pause period ends, the M540
generates acoustic and visual alarm signals as
needed.
To activate alarm monitoring after pausing
1 Press the Alarms function key.
2 Touch All alarms pause before the alarm
pause period ends to cancel the alarm pause.
Alarms
97