Technical data
SpO
measuring accuracy
2
1) 2) 3)
neonatal
PLS measuring accuracy
SpO
/PLS response time
2
Nominal wavelength
Radiant power
NOTE
1)
Since pulse oximeter measurements are statistically distributed, only about two-thirds of those mea-
surements can be expected to fall within ±1 A
2)
The Infinity MCable – Nellcor OxiMax pulse oximeter with adult sensors has been validated in human
blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70 to100 % SpO
against a laboratory co-oximeter and ECG monitor. This variation equals ±1 A
sured by a co-oximeter. Subjects were male (45%) and female (55%) between the ages of 19 and 48,
with skin pigmentation ranging from light to dark.
3)
Accuracy of saturation measurements on neonates is decreased by ±1 digit as compared to
accuracy on adult patients to account for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.
4)
The pulse rate accuracy has been validated on healthy adult volunteers during induced hypoxia stud-
ies in the range of 70 to100 % SpO
equals ±1 A
of the pulse rate value measured by the ECG monitor.
rms
332
0 to 60 % not specified
60 to 80 % not specified: SoftCare SC-PR, SoftCare SC-
NEO, OxiCliq N, D-YS, Oxi-A/N
60 to 80 %, ±3 % for: OxiMax MaxN
70 to 100 %
±2 % for: OxiMax MaxN, SoftCare SC-PR,
SoftCare SC-NEO
±3.5 % for: OxiCliq N
±4 % for: D-YS, Oxi-A/N
4)
PLS:±3 bpm or ±3 % (whichever is greater)
Normal mode: 90 % change within 5 to 7 s
Fast mode: 90 % change within 2 to 4 s
Red: 660 nm
IR: 910 nm
≤15 mW
rms
against a laboratory co-oximeter and ECG monitor. This variation
2
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
of the value measured by a co-oximeter.
2
of the value mea-
rms