Temperature - Dräger Infinity Acute Care System M540 Instructions For Use Manual

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Technical data

Temperature

Parameter display
Measurement range
Resolution
Accuracy (exclusive of probe)
Probe accuracy
Average update time
Response time
Degree of protection against electric
shock
Defibrillation protection
336
Temperatures: Ta, Tb, ΔT, T1a, T1b, ΔT1
Ta, Tb, T1a, T1b: 0 to 50 °C (32 to 122 °F)
ΔT, ΔT1: 0 to 50 °C (0 to 90 °F)
0.1 °C (0.1 °F)
Ta, Tb, T1a, T1b: ±0.1 °C (±0.2 °F)
ΔT, ΔT1: ±0.2 °C (±0.4 °F)
Reusable probe: ± 0.1 °C (±0.2 °F) at 0 °C to 50 °C (32 to 122
°F)
Disposable probes:
± 0.1 °C (±0.2 °F) at 25 °C to 45 °C (77 to 113 °F)
± 0.2 °C (±0.4 °F) at 0 to 25 °C (32 to 77 °F)
<2.5 s
23 to 44 °C (73.4 to 111.2 °F)
±0.2 °C (±0.4 °F) within 150 s
Type CF
Yes
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n