Dräger Infinity Acute Care System M540 Instructions For Use Manual page 263

Patient monitor
Hide thumbs Also See for Infinity Acute Care System M540:
Table of Contents

Advertisement

Priority
Message
None
Offline
None
Recording stored
None
Remote relearn
None
Remote limit change
None
Restore Profile Failed
None
Save profile failed
None
Profile transfer failed
Instructions for use Infinity Acute Care System – Infinity M540 SW VG4.n
Cause
A wireless M540 is un-
docked and does not have
a bed label assignment.
Disconnected network ca-
ble.
A recording was requested
but no recorder is available
for printing the recording.
The indicated function was
initiated from the central
station.
The indicated function was
initiated from the central
station.
The profile could not be re-
stored.
The profile could not be
saved on the M540 and/or
the M500.
The profile failed to load on
the M540 upon docking.
Troubleshooting
Remedy
– Check the network connec-
tions.
– Assign a bed label to the
M540.
Dock the M540 to an the M500 in
an IACS configuration so the re-
cording request can be processed
by the Cockpit.
Informational message – no ac-
tion required.
Informational message – no ac-
tion required.
Try to restore the profile again be-
fore contacting your technical per-
sonnel.
Try to save the profile again be-
fore contacting your technical per-
sonnel.
Undock and redock the M540 be-
fore contacting your technical per-
sonnel.