Format the hard-drive. WARNING: this will erase all data.
FD:
Only use if a new disk has been installed. Note: only
supported if local disk-drive option installed.
Set the length of the recorded files where d is the number of
FLdddd:
pings per file. Default is 5000. Range is from 500 to 50000.
HBd:
Sets the heartbeat message divisor. Heartbeat messages are
output once every 'd' pings where d is a value from 0 to
256. Note that a value of 0 shuts off messages. Default is 0
(off).
Inputs the fish position. The format of the data is
LL<data>:
aaa.aa,N,bbb.bb,E where aaa.aa is the latitude, N is N/S,
bbb.bb is longitude and E is E/W.
MVd.d:
Sets the minimum velocity in m/s. Default is 1.0 m/s.
Opens a new data file regardless of the number of pings
NF:
written to the existing file. This can be used to synchronize
recording to another event such as the beginning of a track.
The current file is closed and the new file is opened on the
subsequent ping following this command. If record mode is
"0" the command will not take effect until record mode is
switched to "1".
Reboot the TPU. In the event that the TPU watchdog reports
RB:
an error the TPU should be rebooted using this command.
Prior to issuing the reboot command recording must be
stopped and all files closed. Refer to status messages for
more information.
Start or stop recording data, where d = 0 stops and d = 1
RCd:
starts recording. Not affected by the set defaults command.
Default (startup condition) is off. Note: only supported if
local disk-drive option installed.
Sets the responder divisor to 2^d, where d = 0 – 15. A value
RDd:
of 15 turns the responder off. Default is 15 (off).
Sets the responder frequency to 24 kHz + d * 500 Hz, where
RFd:
d = 0 – 15. Default is 7 (27.5 kHz).
Set resolution mode where d = 0 is normal and d = 1 is high
Red:
resolution. Note ranges of 100 m and 150 m are not
available in high-resolution mode. Default is normal.
Set range scale, where d = 0 is 50 m, d = 1 is 75 m, d = 2 is
RSd:
100 m and d = 3 is 150 m. Default = 100 m.

Remote Control

B-7