2.9.1.3

CT-PET Emergency Stops

WARNING
ACTIVATION OF THE CT EMERGENCY STOP (E-STOP) DOES NOT STOP MOVEMENT OR
SHUT DOWN POWER TO THE PET MOTION CONTROL SUBASSEMBLY OF THE DLS
SYSTEM. ONLY THE E-STOP BUTTON, LOCATED INSIDE THE PET GANTRY COVERS WILL
STOP THE PET MOTION CONTROL SUBSYSTEM IMMEDIATELY AND SHUT DOWN POWER
TO THE MOTION CONTROL.
Two E-Stops (emergency stop buttons) are provided for the use by service personnel when the
gantry covers are removed for service operations.
• Refer to
6
Figure
2-29. The CT gantry E-Stop button is located on the right side of the gantry
close to the STC back plane.
STC
E-Stop

Figure 2-29: CT Gantry E-Stop