Figure 6-71: Drive Belt Fastener Plate; Figure 6-72: Drive Belt Idler Pulley - GE Discovery LS System Service Manual

Table of Contents

Advertisement

GE H
EALTHCARE
D
2307224-100, R
IRECTION
EVISION
2.) Refer to
over the fastener plate.
3.) Optional: Loosen the idler pulley to release tension on the belt.
a.) Refer to
b.) Loosen the lower right screw, but leave it in place.
Page 162
6
Figure
6-71. Wrap the belt around the two drive pulleys, and bring the ends together
Clamp belt into place
over the fastener plate
(shown with clamp
hardware).

Figure 6-71: Drive Belt Fastener Plate

Figure
6-72. Use a 5 mm Allen wrench to remove the top two and bottom left
socket-head cap screws that fasten the drive belt idler pulley to the side rail.
Four Socket-Head Cap Screws (Right lower screw left in place.)
Drive Belt Idler Pulley

Figure 6-72: Drive Belt Idler Pulley

Chapter 6 - Table
D
LS S
S
ISCOVERY
YSTEM
ERVICE
M
ANUAL