4. Fast Cals (Always After Tube
1.
2.
3.
4.
5.
6
3. Tube Warm-Up
Exam Rx
Daily Prep
Tube Warm-Up
Accept and Run Tube Warm-Up.
Quit
Warm-Up)
Exam Rx
Daily Prep
Fast Cals
OK
Quit
Appendix B - System Calibration Quick Guide
D
ISCOVERY
1 Minute
Do Tube Warm-Up if no CT scans were done in
the last two hours. During activation, a message
box displays in the upper right of the screen. In
addition, information about the process displays
in the center of the screen.
6-30 Minutes
During calibration, a message box displays with
information about the process. If calibration stops
(rare), click [Resume] in the message box. If
calibration does not resume, call Service.
LS S
S
M
YSTEM
ERVICE
ANUAL
Page 349